PGMs
Showing all 21 results
QCRM-7-036PGM Concentrate (Merensky) – 31.8g/t Pt; 16.84g/t Pd; 55.3g/t PGM+Au – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Pd | g/t | 16.84 | ±0.46 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 2.23 | ±0.18 |
Cr2O3 | % | 1.10 | ±0.02 |
Cu | % | 0.47 | ±0.10 |
Ir | g/t | 1.75 | ±0.47 |
Ni | % | 1.03 | ±0.25 |
PGM+Au | g/t | 55.3 | ±2.8 |
Pt | g/t | 31.8 | ±1.0 |
Rh | g/t | 4.25 | ±0.17 |
Ru | g/t | 7.48 | ±0.62 |
QCRM-7-036 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analyzing similar materials.
QCRM-7-066PGM Concentrate
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 2.48 | ±0.50 |
Cr2O3 | % | 1.52 | ±0.10 |
Ni | % | 1.21 | ±0.07 |
Pd | g/t | 21.76 | ±0.88 |
Pt | g/t | 39.8 | ±2.7 |
Rh | g/t | 5.56 | ±0.27 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Co | % | 0.040 | ±0.005 |
Cu | % | 0.56 | ±0.04 |
FeO | % | 14.98 | ±0.85 |
Ir | g/t | 2.04 | ±0.40 |
PGM + Au (Grav) | g/t | 70.5 | ±2.2 |
Ru | g/t | 10.4 | ±1.1 |
QCRM-7-066 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-075PGM Concentrate – 51.1g/t Pt; 17.91g/t Pd; 82.8g/t PGM+Au – Tharisa Minerals
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 0.16 | ±0.02 |
Cr2O3 | % | 1.60 | ±0.08 |
Cu | % | 0.18 | ±0.01 |
FeO | % | 9.47 | ±0.17 |
Pd | g/t | 17.91 | ±0.58 |
PGM+Au (Gravimetric) | g/t | 82.8 | ±4.8 |
Pt | g/t | 51.1 | ±2.2 |
Rh | g/t | 9.23 | ±0.45 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Co | % | 0.024 | ±0.003 |
Ru | g/t | 12.8 | ±1.2 |
QCRM-7-075 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM concentrate and as a calibration standard for the calibration of equipment used for analyzing similar materials.
QCRM-7-079PGM Concentrate (UG2) – 27.2g/t Pt; 12.03g/t Pd; 43.9g/t PGM+Au – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Pd | g/t | 12.03 | ±0.50 |
Pt | g/t | 27.2 | ±1.6 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 1.63 | ±0.27 |
Co | % | 0.03 | ±0.01 |
Cr2O3 | % | 2.50 | ±0.26 |
Cu | % | 0.36 | ±0.06 |
FeO | % | 13.0 | ±2.4 |
Ir | g/t | 1.11 | ±0.31 |
Ni | % | 0.63 | ±0.19 |
PGM+Au | g/t | 43.9 | ±4.4 |
Rh | g/t | 2.88 | 0.24 |
Ru | g/t | 6.12 | ±0.73 |
QCRM-7-079 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analyzing similar materials.
QCRM-7-081PGM Ore (Merensky) – 0.92g/t Pt; 0.42g/t Pd; 1.46g/t PGM+AU – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 0.12 | ±0.02 |
Cr2O3 | % | 2.43 | ±0.13 |
Ni | % | 0.10 | ±0.01 |
Pd | g/t | 0.42 | ±0.02 |
PGM+Au (Grav) | g/t | 1.46 | ±0.07 |
Pt | g/t | 0.92 | ±0.05 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Co | % | 0.011 | ±0.003 |
Cu | % | 0.031 | ±0.014 |
FeO | % | 9.96 | ±0.43 |
Ir | g/t | 0.044 | ±0.017 |
Rh | g/t | 0.086 | ±0.006 |
Ru | g/t | 0.19 | ±0.03 |
QCRM-7-081 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-082PGM Concentrate (Merensky) – 24.96g/t Pt; 10.69g/t Pd; 39.1g/t PGM+Au – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Cr2O3 | % | 2.64 | ±0.05 |
FeO | % | 13.26 | ±0.67 |
Pd | g/t | 10.69 | ±0.54 |
PGM+Au (Grav) | g/t | 39.1 | ±1.7 |
Pt | g/t | 24.96 | ±0.75 |
Rh | g/t | 2.28 | ±0.11 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 1.73 | ±0.11 |
Co | % | 0.025 | ±0.002 |
Cu | % | 0.34 | ±0.03 |
Ir | g/t | 1.00 | ±0.11 |
Ni | % | 0.58 | ±0.04 |
Ru | g/t | 4.79 | ±0.41 |
QCRM-7-082 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-084PGM Ore (Merensky) – 0.79g/t Pt; 0.33g/t Pd; 1.24g/t PGM+AU – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 0.10 | ±0.02 |
Cr2O3 | % | 4.79 | ±0.17 |
Pd | g/t | 0.33 | ±0.02 |
PGM+Au (Grav) | g/t | 1.24 | ±0.09 |
Pt | g/t | 0.79 | ±0.08 |
Rh | g/t | 0.068 | ±0.002 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Co | % | 0.010 | ±0.002 |
Cu | % | 0.018 | ±0.005 |
FeO | % | 10.65 | ±0.54 |
Ir | g/t | 0.03 | ±0.02 |
Ni | % | 0.08 | ±0.01 |
Ru | g/t | 0.16 | ±0.04 |
QCRM-7-084 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-085PGM Concentrate (Merensky) – 30.5g/t Pt; 9.28g/t Pd; 44.6g/t PGM+Au – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 2.69 | ±0.19 |
Cr2O3 | % | 1.38 | ±0.06 |
Pd | g/t | 9.28 | ±0.38 |
PGM+Au (Grav) | g/t | 44.6 | ±1.4 |
Pt | g/t | 30.5 | ±1.3 |
Rh | g/t | 1.60 | ±0.15 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Co | % | 0.018 | ±0.003 |
Cu | % | 0.36 | ±0.03 |
FeO | % | 11.87 | ±0.55 |
Ni | % | 0.45 | ±0.06 |
Ru | g/t | 4.31 | ±0.17 |
QCRM-7-085 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-088PGM Concentrate
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 2.34 | ±0.18 |
Co | % | 0.021 | ±0.002 |
Cr2O3 | % | 1.96 | ±0.06 |
FeO | % | 12.69 | ±0.41 |
Ni | % | 0.52 | ±0.02 |
Pd | g/t | 11.61 | ±0.39 |
Pt | g/t | 33.32 | ±0.80 |
Rh | g/t | 2.58 | ±0.14 |
Ru | g/t | 5.69 | ±0.37 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Ag | g/t | 2.2 | ±1.3 |
Cu | % | 0.38 | ±0.01 |
Ir | g/t | 1.09 | ±0.12 |
PGM + Au (Grav) | g/t | 49.83 | ±0.99 |
S | % | 1.52 | ±0.11 |
QCRM-7-088 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-092PGM Concentrate (UG2) – 80.1g/t Pt; 24.1g/t Pd; 122.1g/t PGM+Au – Barplats Eland Mine
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 0.28 | ±0.06 |
Co | % | 0.016 | ±0.002 |
Cr2O3 | % | 3.40 | ±0.05 |
Cu | % | 0.09 | ±0.01 |
FeO | % | 7.03 | ±0.32 |
Ir | g/t | 4.51 | ±0.61 |
Ni | % | 0.13 | ±0.02 |
Pd | g/t | 24.1 | ±1.1 |
PGM+Au (Grav) | g/t | 122.1 | ±3.3 |
Pt | g/t | 80.1 | ±3.6 |
Rh | g/t | 10.35 | ±0.63 |
Ru | g/t | 14.6 | ±1.2 |
QCRM-7-092 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-102PGM Concentrate
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 0.10 | ±0.02 |
Cr2O3 | % | 5.76 | ±0.29 |
Cu | % | 0.023 | ±0.004 |
FeO | % | 11.46 | ±0.33 |
Ni | % | 0.09 | ±0.01 |
Pd | g/t | 0.44 | ±0.03 |
Pt | g/t | 1.00 | ±0.08 |
Rh | g/t | 0.104 | ±0.009 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Co | % | 0.012 | ±0.004 |
Ir | g/t | 0.042 | ±0.003 |
PGM + Au (Grav) | g/t | 1.60 | ±0.09 |
Ru | g/t | 0.20 | ±0.02 |
S | % | 0.10 | ±0.03 |
QCRM-7-102 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-109PGM Concentrate
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 3.74 | ±0.24 |
Cr2O3 | % | 1.45 | ±0.02 |
Cu | % | 0.54 | ±0.02 |
FeO | % | 13.1 | ±1.0 |
Ir | g/t | 0.80 | ±0.27 |
Ni | % | 0.70 | ±0.04 |
PGM+Au (Grav) | g/t | 59.9 | ±2.0 |
Rh | g/t | 4.59 | ±0.51 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Co | % | 0.024 | ±0.007 |
Pd | g/t | 13.01 | ±0.31 |
Pt | g/t | 40.37 | ±0.89 |
Ru | g/t | 2.35 | ±0.12 |
S | % | 1.86 | ±0.11 |
QCRM-7-109 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-111PGM Ore
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 0.06 | ±0.01 |
FeO | % | 9.97 | ±0.25 |
Pd | g/t | 0.18 | ±0.03 |
Pt | g/t | 0.38 | ±0.06 |
Rh | g/t | 0.04 | ±0.01 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Co | % | 0.011 | ±0.002 |
Cu | % | 0.012 | ±0.002 |
Ir | g/t | 0.022 | ±0.007 |
Ni | % | 0.07 | ±0.02 |
PGM+Au (Grav) | g/t | 0.88 | ±0.33 |
Ru | g/t | 0.10 | ±0.03 |
QCRM-7-111 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-118PGM Ore
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Co | % | 0.015 | ±0.002 |
Cr2O3 | % | 3.50 | ±0.10 |
Cu | % | 0.082 | ±0.009 |
FeO | % | 7.17 | ±0.53 |
Ir | g/t | 4.64 | ±0.82 |
Ni | % | 0.12 | ±0.01 |
Pd | g/t | 24.5 | ±1.6 |
PGM+Au (Grav) | g/t | 120.2 | ±4.1 |
Rh | g/t | 9.7 | ±1.4 |
Ru | g/t | 12.4 | ±2.1 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 0.23 | ±0.05 |
Pt | g/t | 77.4 | ±5.9 |
S | % | 0.54 | ±0.05 |
QCRM-7-118 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-130PGM Tails (UG2) – 0.38g/t Pt; 0.23g/t Pd; 0.66g/t PGM+Au – Sibanye Platinum Mine
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 0.017 | ±0.006 |
Co | % | 0.016 | ±0.001 |
Cr2O3 | % | 12.71 | ±0.13 |
FeO | % | 14.61 | ±0.23 |
Ir | g/t | 0.04 | ±0.01 |
Ni | % | 0.08 | ±0.01 |
Pd | g/t | 0.23 | ±0.02 |
PGM+Au (Grav) | g/t | 0.66 | ±0.05 |
Pt | g/t | 0.38 | ±0.03 |
Rh | g/t | 0.091 | ±0.004 |
Ru | g/t | 0.17 | ±0.01 |
ASSIGNED VALUES (FOR INFORMATION ONLY) | |||
Analytes | Units | Concentrations | Expanded Uncertainty |
Cu | % | 0.006 | ±0.002 |
S | % | 0.05 | ±0.02 |
QCRM-7-130 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Tails and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-134PGM Tails
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 0.017 | ±0.004 |
Co | % | 0.017 | ±0.002 |
Cr2O3 | % | 15.48 | ±0.13 |
FeO | % | 16.43 | ±0.38 |
Ni | % | 0.10 | ±0.01 |
Pd | g/t | 0.26 | ±0.02 |
PGM+Au (Gray) | g/t | 0.62 | ±0.04 |
Pt | g/t | 0.26 | ±0.03 |
Rh | g/t | 0.064 | ±0.008 |
Ru | g/t | 0.146 | ±0.004 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Cu | % | 0.006 | ±0.004 |
Ir | g/t | 0.029 | ±0.004 |
S | % | 0.04 | ±0.03 |
QCRM-7-134 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Tails and as a calibration standard for the calibration of equipment used for analysing similar materials.
QCRM-7-143PGM Tails
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Cr2O3 | % | 11.83 | ±0.31 |
Cu | % | 0.006 | ±0.002 |
Ir | g/t | 0.039 | ±0.005 |
Ni | % | 0.079 | ±0.004 |
Pd | g/t | 0.22 | ±0.02 |
Pt | g/t | 0.36 | ±0.03 |
Rh | g/t | 0.080 | ±0.006 |
Ru | g/t | 0.168 | ±0.011 |
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Co | % | 0.016 | ±0.001 |
FeO | % | 14.32 | ±0.39 |
PGM+Au (Grav) | g/t | 0.60 | ±0.07 |
QCRM-7-143 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Tails and as a calibration standard for the calibration of equipment used for analysing similar materials.
QRM-7-088PGM Concentrate (Merensky) – 33.7g/t Pt; 11.32g/t Pd; 49.4g/t PGM+Au – Metchem Laboratory
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 2.27 | ±0.30 |
Co | % | 0.02 | ±0.01 |
Cr2O3 | % | 1.95 | ±0.08 |
Cu | % | 0.38 | ±0.01 |
FeO | % | 13.0 | ±1.1 |
Ir | g/t | 1.06 | ±0.15 |
Ni | % | 0.52 | ±0.03 |
Pd | g/t | 11.32 | ±0.47 |
PGM+Au | g/t | 49.4 | ±1.3 |
Pt | g/t | 33.7 | ±2.1 |
Rh | g/t | 2.54 | ±0.10 |
Ru | g/t | 5.54 | ±0.43 |
QRM-7-088 is a reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type and verification of analytical methods for PGM Concentrate.
QRM-7-118PGM Concentrate (UG2) – 75g/t Pt; 25.2g/t Pd; 121.3g/t PGM+Au – Barplats Eland Mine
In stock
ASSIGNED VALUES | |||
---|---|---|---|
Analytes | Units | Concentrations | Expanded Uncertainty |
Au | g/t | 0.23 | ±0.05 |
Co | % | 0.017 | ±0.008 |
Cr2O3 | % | 3.43 | ±0.11 |
Cu | % | 0.084 | ±0.015 |
FeO | % | 7.32 | ±0.20 |
Ir | g/t | 5.02 | ±0.47 |
Ni | % | 0.13 | ±0.04 |
Pd | g/t | 25.2 | ±4.0 |
PGM+Au (Grav) | g/t | 121.3 | ±4.1 |
Pt | g/t | 75 | ±11 |
Rh | g/t | 10.8 | ±1.4 |
Ru | g/t | 14.8 | ±1.8 |
QRM-7-118 is a reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type and verification of analytical methods for PGM Concentrate.
QRM-7-122PGM Concentrate
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 3.5 | ±1.5 |
Cr2O3 | % | 1.16 | ±0.10 |
FeO | % | 12.1 | ±3.4 |
Pd | g/t | 10.86 | ±0.41 |
PGM+Au | g/t | 53.55 | ±0.87 |
Pt | g/t | 37.2 | ±1.8 |
Rh | g/t | 1.83 | ±0.29 |
Ru | g/t | 4.0 | ±1.8 |
QRM-7-122 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.
QRM-7-142PGM Concentrate
In stock
CERTIFIED VALUES | |||
ANALYTES | UNITS | CONCENTRATIONS | EXPANDED UNCERTAINTY |
Au | g/t | 5.19 | ±0.66 |
Co | % | 0.057 | ±0.005 |
Cr2O3 | % | 1.40 | ±0.08 |
Cu | % | 1.24 | ±0.11 |
FeO | % | 12.8 | ±1.4 |
Ir | g/t | 13.5 | ±4.6 |
Ni | % | 2.26 | ±0.29 |
Pd | g/t | 223 | ±11 |
PGM+Au (Gravimetric) | g/t | 544 | ±31 |
Pt | g/t | 231 | ±28 |
Rh | g/t | 48.9 | ±7.1 |
Ru | g/t | 60 | ±25 |
QRM-7-142 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.