Showing all 21 results

QCRM-7-036

PGM Concentrate (Merensky) – 31.8g/t Pt; 16.84g/t Pd; 55.3g/t PGM+Au – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Pdg/t16.84±0.46
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t2.23±0.18
Cr2O3%1.10±0.02
Cu%0.47±0.10
Irg/t1.75±0.47
Ni%1.03±0.25
PGM+Aug/t55.3±2.8
Ptg/t31.8±1.0
Rhg/t4.25±0.17
Rug/t7.48±0.62

QCRM-7-036 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analyzing similar materials.

QCRM-7-066

PGM Concentrate

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t2.48±0.50
Cr2O3%1.52±0.10
Ni%1.21±0.07
Pdg/t21.76±0.88
Ptg/t39.8±2.7
Rhg/t5.56±0.27
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Co%0.040±0.005
Cu%0.56±0.04
FeO%14.98±0.85
Irg/t2.04±0.40
PGM + Au (Grav)g/t70.5±2.2
Rug/t10.4±1.1

QCRM-7-066 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-075

PGM Concentrate – 51.1g/t Pt; 17.91g/t Pd; 82.8g/t PGM+Au – Tharisa Minerals

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t0.16±0.02
Cr2O3%1.60±0.08
Cu%0.18±0.01
FeO%9.47±0.17
Pdg/t17.91±0.58
PGM+Au (Gravimetric)g/t82.8±4.8
Ptg/t51.1±2.2
Rhg/t9.23±0.45
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Co%0.024±0.003
Rug/t12.8±1.2

QCRM-7-075 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM concentrate and as a calibration standard for the calibration of equipment used for analyzing similar materials.

QCRM-7-079

PGM Concentrate (UG2) – 27.2g/t Pt; 12.03g/t Pd; 43.9g/t PGM+Au – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Pdg/t12.03±0.50
Ptg/t27.2±1.6
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t1.63±0.27
Co%0.03±0.01
Cr2O3%2.50±0.26
Cu%0.36±0.06
FeO%13.0±2.4
Irg/t1.11±0.31
Ni%0.63±0.19
PGM+Aug/t43.9±4.4
Rhg/t2.880.24
Rug/t6.12±0.73

QCRM-7-079 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analyzing similar materials.

QCRM-7-081

PGM Ore (Merensky) – 0.92g/t Pt; 0.42g/t Pd; 1.46g/t PGM+AU – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t0.12±0.02
Cr2O3%2.43±0.13
Ni%0.10±0.01
Pdg/t0.42±0.02
PGM+Au (Grav)g/t1.46±0.07
Ptg/t0.92±0.05
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Co%0.011±0.003
Cu%0.031±0.014
FeO%9.96±0.43
Irg/t0.044±0.017
Rhg/t0.086±0.006
Rug/t0.19±0.03

QCRM-7-081 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-082

PGM Concentrate (Merensky) – 24.96g/t Pt; 10.69g/t Pd; 39.1g/t PGM+Au – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Cr2O3%2.64±0.05
FeO%13.26±0.67
Pdg/t10.69±0.54
PGM+Au (Grav)g/t39.1±1.7
Ptg/t24.96±0.75
Rhg/t2.28±0.11
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t1.73±0.11
Co%0.025±0.002
Cu%0.34±0.03
Irg/t1.00±0.11
Ni%0.58±0.04
Rug/t4.79±0.41

QCRM-7-082 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-084

PGM Ore (Merensky) – 0.79g/t Pt; 0.33g/t Pd; 1.24g/t PGM+AU – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t0.10±0.02
Cr2O3%4.79±0.17
Pdg/t0.33±0.02
PGM+Au (Grav)g/t1.24±0.09
Ptg/t0.79±0.08
Rhg/t0.068±0.002
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Co%0.010±0.002
Cu%0.018±0.005
FeO%10.65±0.54
Irg/t0.03±0.02
Ni%0.08±0.01
Rug/t0.16±0.04

QCRM-7-084 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-085

PGM Concentrate (Merensky) – 30.5g/t Pt; 9.28g/t Pd; 44.6g/t PGM+Au – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t2.69±0.19
Cr2O3%1.38±0.06
Pdg/t9.28±0.38
PGM+Au (Grav)g/t44.6±1.4
Ptg/t30.5±1.3
Rhg/t1.60±0.15
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Co%0.018±0.003
Cu%0.36±0.03
FeO%11.87±0.55
Ni%0.45±0.06
Rug/t4.31±0.17

QCRM-7-085 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-088

PGM Concentrate

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t2.34±0.18
Co%0.021±0.002
Cr2O3%1.96±0.06
FeO%12.69±0.41
Ni%0.52±0.02
Pdg/t11.61±0.39
Ptg/t33.32±0.80
Rhg/t2.58±0.14
Rug/t5.69±0.37
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Agg/t2.2±1.3
Cu%0.38±0.01
Irg/t1.09±0.12
PGM + Au (Grav)g/t49.83±0.99
S%1.52±0.11

QCRM-7-088 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-092

PGM Concentrate (UG2) – 80.1g/t Pt; 24.1g/t Pd; 122.1g/t PGM+Au – Barplats Eland Mine

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t0.28±0.06
Co%0.016±0.002
Cr2O3%3.40±0.05
Cu%0.09±0.01
FeO%7.03±0.32
Irg/t4.51±0.61
Ni%0.13±0.02
Pdg/t24.1±1.1
PGM+Au (Grav)g/t122.1±3.3
Ptg/t80.1±3.6
Rhg/t10.35±0.63
Rug/t14.6±1.2

QCRM-7-092 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-102

PGM Concentrate

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t0.10±0.02
Cr2O3%5.76±0.29
Cu%0.023±0.004
FeO%11.46±0.33
Ni%0.09±0.01
Pdg/t0.44±0.03
Ptg/t1.00±0.08
Rhg/t0.104±0.009
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Co%0.012±0.004
Irg/t0.042±0.003
PGM + Au (Grav)g/t1.60±0.09
Rug/t0.20±0.02
S%0.10±0.03

QCRM-7-102 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-109

PGM Concentrate

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t3.74±0.24
Cr2O3%1.45±0.02
Cu%0.54±0.02
FeO%13.1±1.0
Irg/t0.80±0.27
Ni%0.70±0.04
PGM+Au (Grav)g/t59.9±2.0
Rhg/t4.59±0.51
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Co%0.024±0.007
Pdg/t13.01±0.31
Ptg/t40.37±0.89
Rug/t2.35±0.12
S%1.86±0.11

QCRM-7-109 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-111

PGM Ore

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t0.06±0.01
FeO%9.97±0.25
Pdg/t0.18±0.03
Ptg/t0.38±0.06
Rhg/t0.04±0.01
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Co%0.011±0.002
Cu%0.012±0.002
Irg/t0.022±0.007
Ni%0.07±0.02
PGM+Au (Grav)g/t0.88±0.33
Rug/t0.10±0.03

QCRM-7-111 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Ore and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-118

PGM Ore

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Co%0.015±0.002
Cr2O3%3.50±0.10
Cu%0.082±0.009
FeO%7.17±0.53
Irg/t4.64±0.82
Ni%0.12±0.01
Pdg/t24.5±1.6
PGM+Au (Grav)g/t120.2±4.1
Rhg/t9.7±1.4
Rug/t12.4±2.1
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t0.23±0.05
Ptg/t77.4±5.9
S%0.54±0.05

QCRM-7-118 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-130

PGM Tails (UG2) – 0.38g/t Pt; 0.23g/t Pd; 0.66g/t PGM+Au – Sibanye Platinum Mine

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t0.017±0.006
Co%0.016±0.001
Cr2O3%12.71±0.13
FeO%14.61±0.23
Irg/t0.04±0.01
Ni%0.08±0.01
Pdg/t0.23±0.02
PGM+Au (Grav)g/t0.66±0.05
Ptg/t0.38±0.03
Rhg/t0.091±0.004
Rug/t0.17±0.01
ASSIGNED VALUES (FOR INFORMATION ONLY)
AnalytesUnitsConcentrationsExpanded Uncertainty
Cu%0.006±0.002
S%0.05±0.02

QCRM-7-130 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Tails and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-134

PGM Tails

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t0.017±0.004
Co%0.017±0.002
Cr2O3%15.48±0.13
FeO%16.43±0.38
Ni%0.10±0.01
Pdg/t0.26±0.02
PGM+Au (Gray)g/t0.62±0.04
Ptg/t0.26±0.03
Rhg/t0.064±0.008
Rug/t0.146±0.004
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Cu%0.006±0.004
Irg/t0.029±0.004
S%0.04±0.03

QCRM-7-134 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Tails and as a calibration standard for the calibration of equipment used for analysing similar materials.

QCRM-7-143

PGM Tails

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Cr2O3%11.83±0.31
Cu%0.006±0.002
Irg/t0.039±0.005
Ni%0.079±0.004
Pdg/t0.22±0.02
Ptg/t0.36±0.03
Rhg/t0.080±0.006
Rug/t0.168±0.011
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Co%0.016±0.001
FeO%14.32±0.39
PGM+Au (Grav)g/t0.60±0.07

QCRM-7-143 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QCRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Tails and as a calibration standard for the calibration of equipment used for analysing similar materials.

QRM-7-088

PGM Concentrate (Merensky) – 33.7g/t Pt; 11.32g/t Pd; 49.4g/t PGM+Au – Metchem Laboratory

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t2.27±0.30
Co%0.02±0.01
Cr2O3%1.95±0.08
Cu%0.38±0.01
FeO%13.0±1.1
Irg/t1.06±0.15
Ni%0.52±0.03
Pdg/t11.32±0.47
PGM+Aug/t49.4±1.3
Ptg/t33.7±2.1
Rhg/t2.54±0.10
Rug/t5.54±0.43

QRM-7-088 is a reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type and verification of analytical methods for PGM Concentrate.

QRM-7-118

PGM Concentrate (UG2) – 75g/t Pt; 25.2g/t Pd; 121.3g/t PGM+Au – Barplats Eland Mine

In stock

ASSIGNED VALUES
AnalytesUnitsConcentrationsExpanded Uncertainty
Aug/t0.23±0.05
Co%0.017±0.008
Cr2O3%3.43±0.11
Cu%0.084±0.015
FeO%7.32±0.20
Irg/t5.02±0.47
Ni%0.13±0.04
Pdg/t25.2±4.0
PGM+Au (Grav)g/t121.3±4.1
Ptg/t75±11
Rhg/t10.8±1.4
Rug/t14.8±1.8

QRM-7-118 is a reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type and verification of analytical methods for PGM Concentrate.

QRM-7-122

PGM Concentrate

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t3.5±1.5
Cr2O3%1.16±0.10
FeO%12.1±3.4
Pdg/t10.86±0.41
PGM+Aug/t53.55±0.87
Ptg/t37.2±1.8
Rhg/t1.83±0.29
Rug/t4.0±1.8

QRM-7-122 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.

QRM-7-142

PGM Concentrate

In stock

CERTIFIED VALUES
ANALYTESUNITSCONCENTRATIONSEXPANDED UNCERTAINTY
Aug/t5.19±0.66
Co%0.057±0.005
Cr2O3%1.40±0.08
Cu%1.24±0.11
FeO%12.8±1.4
Irg/t13.5±4.6
Ni%2.26±0.29
Pdg/t223±11
PGM+Au (Gravimetric)g/t544±31
Ptg/t231±28
Rhg/t48.9±7.1
Rug/t60±25

QRM-7-142 is a certified reference material which is suitable for use as random control samples in routine analytical laboratory quality control, when inserted within a batch of samples and measured in parallel to the unknown. The QRM can also be used as a control sample in the analysis of samples of a similar type, verification of analytical methods for PGM Concentrate and as a calibration standard for the calibration of equipment used for analysing similar materials.